22 Check list for fault rectification

Mains switch switched on:

no indication on the display of

the temperature module

equipment fuse has blown; 15A
instrument fuse has blown; T100mA 250V~ on board 55162.x
controller faulty (board 55616.x)
power supply interrupted

no fan operation

instrument fuse; T50mA on power supply for internal fan
55145.x has blown

Symbol nor alight

ambient temperature too high
temperature inside the unit higher than the selected setpoint
temperature

Symbol is alight

temperature protection (TB) has been activated

Symbol is flashing

temperature protection (TWW, ASF) has been activated

• and

rh empty

water container empty

• and

rh over

humidity setpoint has been exceeded

E-0

in temperature module

error on self-test

E-1

in temperature module

power unit triac faulty

E-2

in temperature module

power unit faulty

E-3

in temperature module

Pt100 temperature sensor faulty

E-6

in humidity module

error sensor rh

humidity sensor faulty

E-L1

in timer module

error communication to power unit L1

As far as HCP appliances are concerned, error messages are shown in the alphanumeric display.
In case there is a fault on the oven, please get in touch with an authorised service organisation or contact
the MEMMERT customer service department. (see section 25)
In case of queries always specify model and serial number (on the rating label)

23 Supply failure