Loadstar Sensors DS-3000 Digital Load Cell Display & Controller User Manual
Page 42

DS-3000U
DS3000 User Guide
Loadstar Sensors, Inc.
Page: 42 / 48
033-01419 Rev. 4
11.4 CT0 ©
Function:
TARE all currently connected sensors simultaneously
Summary:
This remote command tares all connected sensors at the same time.
11.5 ETT © (Future Command, not yet implemented)
Function:
Set function of external input
Keystrokes:
These commands remotely control the options mode page trgr entry.
ETT D
©
ETT T
©
ETT R
©
ETT B
©
Remote equivalent to Trgr
Remote equivalent to trgr t.
Remote equivalent to trgr r
Remote equivalent to trgr b
Summary:
Configure the function of external input.
11.6 TTS, TTH ©
Function:
Set tare type
Keystrokes:
These commands tell the DS-3000 how to handle default tare requests.
TTS
©
TTH
©
Soft tare attached sensors, equivalent to programming mode
page tare, field type s.
Hard tare attached sensors, equivalent to programming
mode page tare, field type h
Summary:
Tell the DS-3000 how to handle user tare requests (both front panel and
remote).
11.7 SST ©
Function:
Read Sensors internal status code
Keystrokes:
This is a debug command.
SST
©
Returns internal status code.
Summary:
Returns internal status code. This is a read only value.
11.8 LED© (Future Command, Not yet Implemented)
Function:
Sets intensity of display LED
Keystrokes:
This command sets the brightness of the display LED. The range is from 10 to