Loadstar Sensors DS-3000 Digital Load Cell Display & Controller User Manual
Page 24
![background image](/manuals/566397/24/background.png)
DS-3000U
DS3000 User Guide
Loadstar Sensors, Inc.
Page: 24 / 48
033-01419 Rev. 4
Summary:
Select which sensor, or combination of sensors is currently displayed on the
front panel LED display. Does not affect the value returned by the USB uplink
port to a connected PC (all four summed).
8.2.3
Tare Options Menu
Function:
TARE Connected Sensors
Keystrokes:
Mode Key menu till tare is displayed, then press up or down to select input
sensor to tare, then press enter.
tare 1 ©
tare 2 ©
tare 3 ©
tare 4 ©
type
Tare connected Sensor 1
Tare connected Sensor 2
Tare connected Sensor 3
Tare connected Sensor 4
NOTE: Tare all connected Sensors simultaneously, is done using
enter
key in run mode.
Press enter to toggle between type h and type s. Most
applications should use type h (or “hard tare”). If you have an
application that requires very frequent tares (>10/hour), type s (or
“soft tare”) may be more appropriate. Please contact Loadstar
Sensors support for more information.
Summary:
Manually Tare individual sensors. Set the type of tare command sent to the
sensors.
8.2.4
Relay Output 1 Menu
Function:
Setup output contact 1 related parameters
Keystrokes:
Press menu till reLay1 is displayed, then press up or down to select which
parameter you’d like to adjust, then press enter to edit each entry to the desired
value.
These parameters affect Relay Output 1 only! Relay 1 and Relay 2 are
independent, and are programmed separately.
trgr 1
First trigger action threshold. Select enter to enter the
numeric entry field. up and down changes the value of the
currently highlighted digit, and menu changes the currently
highlighted digit. Press enter when finished
Pnt1
pnt1ta Activate output 1 above this value, ie. “Trigger Above”
pnt1tB Activate output 1 below this value, ie. “Trigger Below”
pnt1ra Reset output 1 above this value, ie. “Reset Above”
pnt1rb Reset output 1 below this value, ie. “Reset Below”
Trgr 2
Second hysteresis action threshold. . Select enter to enter
the numeric entry field. up and down changes the value of
the currently highlighted digit, and menu changes the