6. Press
enter
to tare the selected sensor. The chosen port decimal point will flash when taring is
completed. You may individually tare any sensors you like in this manner.
7. Exit programming mode when completed: Press mode to the run entry.
8. Then
press enter to return to run mode. The display will now reflect the new result based on the
newly tared sensor.
7.5 How can I tell which sensors are currently being displayed?
1. Press and hold menu until “Enter Pass” and “000000” appear on the display
2. Press
the
up
or down keys to select the programming mode password (default is 000001)
3. Press
enter
4. You are now in programming mode at the dsplay prompt. Notice the decimal points are flashing
for inactive ports, and solid on for active ports. Thus any port with a solid decimal point is
summed in the current display mode. Any flashing port is enabled to be summed, but is
currently inactive and not summed/displayed because no sensor is currently installed in that port
position.
5. To exit back to run mode: Press mode until the run entry.
6. Then
press enter to return to run mode.
7.6 How do I get the firmware version?
1. Press and hold menu until “Enter Pass” and “000000” appear on the display
2. Press
the
up
or down keys to select the programming mode password (default is 000001)
3. Press
enter
4. You are now in programming mode at the dsplay prompt. Press mode until you reach the
option menu.
5. Press
down
, three times to the “v3 05” this is the current firmware revision. At the time of
this writing, version 3.05
6. Now exit programming mode: Press mode until the run entry
7. Then
press enter to return to run mode. Closing the remote contacts will now automatically
tare sensors, same as the onscreen tare pushbutton.
7.7 How do I enable the remote tare function?
1. Press and hold menu until “Enter Pass” and “000000” appear on the display
2. Press
the
up
or down keys to select the programming mode password (default is 000001)
3. Press
enter
4. You are now in programming mode at the dsplay prompt. Press mode until you reach the
option menu.
5. Press
down
, to the trgt entry. Press enter until trgt t is displayed.
6. Now exit programming mode: Press mode until the run entry
7. Then
press enter to return to run mode. Closing the remote contacts will now automatically
tare sensors, same as the front panel enter pushbutton.
7.8 How do I change the display brightness?
1. Press and hold menu until “Enter Pass” and “000000” appear on the display