Loadstar Sensors DS-3000 Digital Load Cell Display & Controller User Manual
Page 19

DS-3000U
DS3000 User Guide
Loadstar Sensors, Inc.
Page: 19 / 48
033-01419 Rev. 4
2. Press
the
up
or down keys to select the programming mode password (default is 000001)
3. Press
enter
4. You are now in programming mode at the dsplay prompt. Press mode until you reach the
option menu.
5. Press
down
, to the LED entry. Press enter and a numeric field appears onscreen.
6. Press
down
and up until the desired brightness is achieved. Maximum brightness is 128
and minimum is 10. Default is 64. If power consumption is a concern, the power
requirements double from 64 to 128, so it is recommended that you choose the minimum
value which is acceptable for your application. Press enter to retain your final setting. The
display changes immediately when this value is edited, so it is easy to determine the best
value.
7. Now exit programming mode: Press mode until the run entry
8. Then
press enter to return to run mode. The display will now retain your newly set display
brightness, even when powered off then powered back on.
7.9 How do I turn off attached sensor ports to save power?
1. Press and hold menu until “Enter Pass” and “000000” appear on the display
2. Press
the
up
or down keys to select the programming mode password (default is 000001)
3. Press
enter
4. You are now in programming mode at the dsplay prompt. Press mode until you reach the
power menu.
5. Press
down
, to the s1 entry. Press enter to toggle this port on or off. S1 corresponds to
port 1, S2 corresponds to port 2, S3 to port 3, and S4 to port 4.
6. Now exit programming mode: Press mode until the run entry
7. Then
press enter to return to run mode.
7.10 How do I change the password?
8. Press and hold menu until “Enter Pass” and “000000” appear on the display
9. Press
the
up
or down keys to select the programming mode password (default is 000001)
10. Press enter
11. You are now in programming mode at the dsplay prompt. Press mode until you reach the
option menu.
12. Press down, to the pass entry. Press enter to toggle to the numeric entry field.
13. Press up and down pushbuttons to increment and decrement the currently selected
(flashing) digit. Press the menu button to highlight the next digit. Repeat as necessary until
your desired password is displayed onscreen.
14. Press enter when you are done. The new password is automatically updated.
15. Now exit programming mode: Press mode until the run entry
16. Then press enter to return to run mode.
NOTE: You will need to use this new password every time you enter
programming mode. Do not lose it! There is no way to reset the unit to a
known password without it, BEWARE !!