Set Relay 2, Trigger Point2 to “Trigger Above.”
Summary:
Remotely set Trigger Type for each programmable set-point.
11.12 RnSx© (Future Command, Not yet implemented)
Function:
Sets sensor x relationship to Relay n
Keystrokes:
This command sets which sensors are associated with which relays. For
example, Relay one may be used with one sensor, and Relay two may be used
with a pair of others. Options are SUM “S”(Relay n reacts to the SUM of all
associated inputs), INDIVIDUAL “I”, (Relay n reacts only to this individual
sensor),or NONE “N” (this relay does not respond to this sensor)
R1S1 S ©
R1S2 S©
R1S3 N©
R1S4 N©
R2S1 S©
R2S2 S©
R2S3 S©
R2S4 S©
Set Relay 1 trigger points to sum with sensor 1
Set Relay 1 trigger points to sum with sensor 2
Set Relay 1 trigger points to ignore sensor 3
Set Relay 1 trigger points to ignore sensor 4
Set Relay 2 trigger points to sum with sensor 1
Set Relay 2 trigger points to sum with sensor 2
Set Relay 2 trigger points to sum with sensor 3
Set Relay 2 trigger points to sum with sensor 4
Summary:
Set which sensor is used to determine the threshold conditions
11.13 RnD© (Future Command, Not yet Implemented)
Function:
Relay n duration
Keystrokes:
This command sets the duration of the output contact, once a trigger condition is
detected. The duration is in seconds, and 0.0 seconds = continuous. Default is
continuous.
R1D 0.0 ©
R2D 0.0©
Set Relay1 duration to continuous.
Set Relay2 duration to continuous.
Summary:
Set output contact duration.
11.14 ASx©, RSx © (Future Command, Not yet implemented)
Function:
Adds or Removes sensor from current display sum
Keystrokes:
Add or Remove Sensor x sum from the current display.
AS1 ©
AS2©
AS3©
AS4©
Display sum includes sensor 1
Display sum includes sensor 2
Display sum includes sensor 3
Display sum includes sensor 4