Safety – Leica Biosystems RM2265 User Manual
Page 9
7
Leica RM2265
Personal safety precautions
2. Safety
Warnings – Safety notes on the instrument itself
Warnings – Transport and installation
• Safety notes on the instrument itself, which are marked with a warning triangle, indicate that
the correct operating instructions (as defined in these Instructions for Use) must be followed
when operating or replacing the item marked.
• Failure to adhere to these instructions may result in an accident, personal injury, damage to
the instrument or accessory equipment.
• Once unpacked, the instrument may be transported only in an upright position.
• Never lift the instrument by the handwheels or the cassette clamp. Always remove the sec-
tion waste tray before transporting the instrument.
• Caution! The voltage selector has been preset at the factory.
• Before connecting the instrument to the line voltage, please check that this setting complies
with the local power requirements of your laboratory.
• The mains cable inlet is closed with an adhesive tape that indicates the actual voltage set-
ting of the instrument upon arrival. Severe damage can be caused to the instrument if the
voltage selector is set to an incorrect voltage!
• When adjusting the voltage selector setting, the instrument must be disconnected from the
mains.
• The instrument must be connected to mains only with one of the supplied power cords and
only to grounded sockets. Do not interfere with the grounding function by using an extension
cord without a ground wire.
• The instrument is ready for operation only if a dummy plug or foot switch (optional) is con-
nected. If you hear a beeping sound (continuous) after switching on the instrument, check
that the control panels are connected correctly.
• The instrument may not be operated in hazardous locations.
• Extreme temperature fluctuations between storage facility and setup site as well as high
humidity may cause condensation to form. If this is the case, wait at least two hours before
switching on the instrument. Failure to comply with this may cause damage to the instrument.
• Protective devices on both instrument and accessories must neither be removed nor modified.
• When working with microtomes, personal safety precautions must always be taken. It is
mandatory to wear work safety shoes, safety gloves, a mask and safety goggles.