Instrument setup – Leica Biosystems RM2265 User Manual
Page 21
19
Leica RM2265
4. Instrument Setup
22
25
24
21
23
The voltage selector is located above the main
power switch, on the left-rear side of the instru-
ment (
Fig. 12
). The voltage setting is displayed in
the viewing window (22).
• Insert a small screwdriver into the cutout (25)
and carefully pry out the insert.
• Remove the voltage selector housing (21)
together with the fuses (23).
Remove the voltage selector block (24) (white)
and insert it again such that the correct local
voltage is displayed in the viewing window
(22).
• Reinsert the voltage selector housing with the
block and fuses and push it in until it engages
(audible click).
21
26
Connecting the power supply
• Before connecting the power cable, make sure that the power switch
(27) on the rear of the instrument is switched to "O" = OFF.
• Various country-specific power cables are provided with the instrument.
Make sure that the power cable used has the correct plug for the power
socket.
• Insert the connector of the power cable into the connection socket (26)
and plug the power plug into the power socket.
27
Checking the voltage
(continued)
Exposure to extreme temperature changes and high air humidity may cause condensa-
tion to form inside the instrument.
After transporting, wait at least 2 hours to allow the instrument to adopt the ambient
temperature before turning it on!
Failure to comply with this may cause damage to the instrument.
Fig. 12