Troubleshooting – Leica Biosystems RM2265 User Manual
Page 77
75
Leica RM2265
Problem
Possible cause
Corrective action
4. Noises during sectioning
The knife "sings" when section-
ing hard specimens. The sec-
tions show scratches or chatter
marks.
• The sectioning speed is too fast.
• The clearance angle is too
wide.
• Insufficient clamping of speci-
men clamp and/or knife holder.
• Select a slower sectioning
speed.
• Methodically reduce the clear-
ance angle setting until you have
found the optimum angle.
• Check all screw and clamp con-
nections on the specimen holder
system and the knife holder. If
necessary, tighten the levers and
screws.
5. High blade consumption
• Too great of a sectioning force
was applied.
• Adjust the sectioning speed and/
or section thickness when sec-
tioning. Select a thinner section.
7. Troubleshooting
- SM2010 R (46 pages)
- CM305 S (66 pages)
- CE Knife holder (14 pages)
- CM1950 (78 pages)
- VT1200/VT1200 S (62 pages)
- VT1000 S (42 pages)
- HI1220 (18 pages)
- HI1210 (18 pages)
- CM1850 UV (56 pages)
- CM1850 UV (54 pages)
- CM1510-S (56 pages)
- CM3600 XP (114 pages)
- CM1100 (36 pages)
- CM1860/CM1860 UV (98 pages)
- Premium blade holder (14 pages)
- CM1520 (92 pages)
- VT1000 A (49 pages)
- VT1000 P (48 pages)
- SP2600 (51 pages)
- SM2500 (76 pages)
- SM2400 (8 pages)
- SM2000R (22 pages)
- EG1160 (32 pages)
- EG1150 C (20 pages)
- EG1150 H (42 pages)
- EG1130 (2 pages)
- EG1120 (2 pages)
- EG F (18 pages)
- ASP300 (70 pages)
- PELORIS_PELORIS II (160 pages)
- TP1020 (102 pages)
- ASP200 S (88 pages)
- ASP300 S (88 pages)
- ASP6025 (178 pages)
- LN22 (44 pages)
- RM CoolClamp (20 pages)
- RM2255 (82 pages)
- RM2245 (78 pages)
- RM2235 (78 pages)
- RM2125 RTS (60 pages)
- SP9000 (30 pages)
- SP1600 (10 pages)
- ST4040 (50 pages)
- ST5020 (88 pages)