2 wartungshinweise, 2 maintenance instructions, Cleaning and maintenance – Leica Biosystems RM2265 User Manual
Page 81

79
Leica RM2265
8. Cleaning and Maintenance
8.2.2 Maintenance instructions
Note for inspection
• When the instrument reaches approx. 2 million performed strokes, the
code SEr always appears for approx. 3 seconds in the three-digit display
when the TRIM/SECT button is pressed
(switching between trimming and sectioning mode).
This is a reminder that the instrument must be inspected by an autho-
rized Leica service technician, regardless of how heavily the instrument
has been used.
↓
The instrument is basically maintenance-free.
To ensure trouble-free operation of the instrument over a long period of time, the following is recom-
mended by Leica:
3. Have the instrument inspected on a regular
basis by a qualified service technician autho-
rized by Leica.
The intervals depend on how heavily the in-
strument is used.
1. Thoroughly clean the instrument on a daily
basis.
2. Enter into a service contract at the latest
at the end of the warranty period. For more
information, please contact your local Leica
technical service center.
We recommend the following inspection intervals depending on the general workload of the instru-
ment as defined in the two categories listed in the table:
Only authorized and qualified service personnel may access the internal components of
the instrument for service and repair!
Category I
Category II
Number of sections per day:
> 8,000 sections per day
< 8,000 sections per day
Workload (hours per day):
> 5 hours per day
<
5 hours per day
Sectioning speed:
Predominantly high sectioning
speed
Slow to medium sectioning speed
Specimen material:
Working with soft and hard speci-
mens
Predominantly soft specimen
material
Maintenance:
Every 12 months
Every 24 months