Optional accessories – Leica Biosystems RM2265 User Manual
Page 68
66
Instructions for Use, V 1.8 RevC – 07/2013
2. Attach vertical column
• Insert the cap screw (b) into the hole of the
baseplate from below. Place the lock washer
(c) on the cap screw from above.
• Thread the silver vertical column (2) onto the
baseplate from above and tighten with the
Allen key size 8.
A
3. Attach horizontal arm
• Slide the support ring (e) onto the vertical
column and position it so that the lock nut (f)
is facing to the back. Tighten the lock nut.
• Slide the cross-member (d) onto the column.
Ensure that the lock screw (g) is facing to the
right of the baseplate. The horizontal arm must
be centered over the microscope.
• Slide the horizontal arm (3), flat side facing the
lock screw (g), into the cross-member (d) and
tighten.
6.
Optional Accessories
Fig. 64
Fig. 63
Fig. 62
b
c
2
e
d
2
g
f
h
Extreme danger of tipping! After in-
stalling the vertical column, imme-
diately place the microtome onto the
baseplate so that the front feet of the
microtome are positioned in the shal-
low recesses (A).
The lock washer must be positioned
between the baseplate and vertical
column to prevent unintentional turn-
ing of the column.
For more information on connecting
and using the microscope, magnifying
lens or cold-light source, please see
the appropriate Instructions for Use.