Cleaning, disinfection, maintenance – Leica Biosystems CM1860/CM1860 UV User Manual
Page 67
67
Leica CM1860/CM1860UV – Cryostat
Removal of the lamp
The UVC lamp (3) is installed in front
of the protection screen for chamber
illumination (4).
• Hold the lamp with both hands and
carefully pull it out of the clips (5)
with a slight forward movement.
• Detach the metal ring (7) on the
holder (6) in the direction of the ar-
row (8) and carefully pull the lamp
out of the holder with your right
hand (
see Fig. 60 + 61
).
Installation of the new lamp
• Carefully slide the metal ring (7) over
the lamp from the left (
see Fig. 62
).
• Push the lamp into the holder on the
left side until it engages.
• Screw the metal ring onto the
holder, then hold the lamp with both
hands and carefully push it into the
clips (5).
• Replace the sliding window.
• Connect the instrument to the power
supply again and switch it on.
4
3
5
7
Fig. 60
Fig. 61
Fig. 62
6
8
Dispose of the UVC lamp separately!
9. Cleaning, Disinfection, Maintenance
If any metallic mercury is
released, handle it carefully
and dispose of it properly!