2 stationary heat extractor (optional), 2 specimen discs, 2 specimen orientation – Leica Biosystems CM1860/CM1860 UV User Manual
Page 36: 2 stationary heat extractor, Optional), Specimen discs, Daily use of the instrument
36
Instructions for Use V1.4 – 10/2012
7.
Daily Use of the Instrument
7.2.1 Inserting the specimen discs into the
specimen head
7.2 Specimen discs
Fig. 25
• Lock the handle of the handwheel in the upper
position.
• If the knife holder and a knife are in place,
cover the knife edge with the knife guard. If
using a CE disposable blade holder, the anti-
roll guide should be lowered to cover the blade
edge.
• Loosen the clamping screw (1,
Fig. 25
) on the
specimen head.
• Insert the shaft of the specimen disc (3) in the
location hole (2) of the specimen head.
Make sure that the shaft of the speci-
men disc is fully inserted. The entire
surface of the rear side of the speci-
men disc must be free of residue.
• Tighten screw (1).
7.2.2 Specimen orientation
• Loosen the clamping lever (4).
• Orient the specimen on the ball shell using the
lever (5).
• Tighten the clamping lever (4).
7.1.2 Stationary heat extractor
(optional)
• Fasten the support (1,
Fig. 24
) of the heat extractor (2,
Fig. 24
) to the left
side panel of the cryochamber using two screws in the provided holes.
Insert the heat extractor.
• Lower the heat extractor cylinder on the specimen surface. After ap-
proximately 30 seconds, the specimen should be completely frozen.
• After the specimen is frozen, move the heat extractor into resting position
at the parking station (3,
Fig. 24
).
1
3
2
4
5
The specimen discs are to be handled
by their side O-rings to protect against
frostbite!