Technical data – Leica Biosystems CM1860/CM1860 UV User Manual
Page 16
16
Instructions for Use V1.4 – 10/2012
3.
Technical Data
Operating temperature range (ambient temperature): 18 °C to 35 °C. All specifications related to
temperature of the cooling unit are valid only for an ambient temperature of 22 °C and a relative
humidity of no more than 60 %!
Type
CM1860/ CM1860/ CM1860/ CM1860/ CM1860/
CM1860 UV
CM1860 UV
CM1860 UV
CM1860 UV
CM1860 UV
Test symbol
-
c_CSA_us
-
-
-
Nominal voltage
100 VAC
120-127 VAC
220-230 VAC
220-230 VAC
240 VAC
±10 %
±10 %
±10 %
±10 %
±10 %
Nominal frequency
50/60 Hz
60 Hz
60 Hz
50 Hz
50 Hz
Power input
1440 VA
1440 VA
1440 VA
1440 VA
1440 VA
Max. start-up current for 5 sec 45 A eff.
45 A eff.
45 A eff.
45 A eff.
45 A eff.
Protective class
I
I
I
I
I
Pollution degree
2
2
2
2
2
Overvoltage inst. category II
II
II
II
II
Heat emission (max.)
1600 J/s
1600 J/s
1600 J/s
1600 J/s
1600 J/s
Refrigeration system
50 Hz/60Hz
Cryochamber
Temperature range
0 °C to -35 °C (+ 3 K/- 3 K)
Cooling time down to -35 °C
max. 6 hrs., at 22 °C ambient temperature
Defrost
Automatic hot gas defrost
1 automatic defrost cycle/24 hours, time-controlled (duration 12 min.)
Refrigeration capacity
690 W
Safety factor
3
Refrigerant
300g ±5g refrigerant R 404A *
Compressor oil
0.6 l EMKARATE RL-22S, ICI *
Quick-freeze shelf
Maximum temperature
- 40 °C (+ 3 K/- 5 K)
Number of freezing stations
8
Defrost
Manual hot gas defrost, time-controlled (duration 12 min.)
Peltier element
Max. temperature difference
17K, at - 35 °C chamber temperature
Number of freezing stations
2
Defrost
In conjunction with the
quick-freeze shelf
according to IEC-1010, UL 3101