4 display lock, 5 control panel 2 - electric coarse feed, Display lock – Leica Biosystems CM1860/CM1860 UV User Manual
Page 34: Control panel 2 - electric coarse feed, Operating the instrument
34
Instructions for Use V1.4 – 10/2012
Advancing the specimen toward the knife
• Starts fast or slow feed toward the knife. The LED (2) starts flashing as
long as the specimen head is in motion.
The LED (2) lights up when the forward end position has been reached.
This is a safety feature designed to protect both the specimen and
the knife.
• To feed the specimen, press or hold the appropriate SLOW or RAPID
button.
6.5 Control panel 2 - Electric coarse feed
Move the specimen away from the knife
• Starts fast retraction to the rear end position.
The LED (1) flashes while the specimen head is in motion.
The LED (1) lights up when the rear end position has been reached.
• The return movement can be stopped by pressing one of the coarse feed
buttons.
• Starts slow retraction to the rear end position.
The motion continues as long as the button is held.
6.
Operating the Instrument
1
2
slow
rapid
Fig. 23
slow
rapid
Fig. 22
6.4 Display lock
After the display is locked using the key button (press and hold for 5 sec.),
set values can no longer be changed.
Only the UV disinfection can still be enabled/disabled by pressing the UV key.
Push the key button once more for 5 seconds to unlock the display.
When the display is locked, the LEDs between the hour and minute indication
on the time panel are turned off.