5. EXECUTING METHODS
Methods defined with Adagio can be executed within the configured
pump system.
To do that, a floating window called Method Execution is provided
through the Execute Methods button located in the experimentation
assistant bar.
The available pump list table shows the pumps already defined within
the pump system together with the method to be executed within each of
them.
Each row of the table shows the following information:
• Method: the name of the method which will be executed within the
pump (see chapter 4 for more details).
• Pump name: the name assigned to the pump during its creation /
edit process (see 3.1.1 for more details).
• Current flow: the real flow infused / withdrawn by the pump
currently. It is shown with 3 decimal places and an I/W character
showing whether a “infusion” (I) or a “withdrawal” (W) is being done
(see 5.2.2 for more details).
• Elapsed time: the time (in hh:mm:ss format) lapsed from the
beginning of the method execution (see 5.2.2 for more details).
• Progress: a bar showing the percentage of the method which has
already been executed. It is calculated as [Current time] / [Duration].
• Duration: the total duration of the method.
• Status: the current execution status of the pump (see 5.2.3 for more
details).
• Individual Pump Control - Start / Stop / Reset: a set of buttons
which allow you to start / pause / resume / stop / reset the Method
execution on each pump independently (see 5.2.1 for more details).