When adding a new step the Step Editor window will automatically
open, allowing you to edit the step’s target values (duration, flow values and
step linkage property). This window can also be accessed by clicking on the
Properties icon in Method Setup or by right-clicking on a step and selecting
‘Properties…’
Changing the flow may affect the duration or volume of the step.
Therefore, the following rules will be applied when the step volume, start /
end flow rates or the duration are changed:
• If the volume changes, then the duration/time will be updated.
• If the duration/time changes, then the volume will be updated.
• If the start/end flow changes, then the volume will be updated.
Additional editable information is available in the Accumulated values
tab page of the Properties window:
• Volume: accumulated real
volume infused / withdrawn by
the method until the end of the
current step. If the value is
changed, then the step volume
(and thus the step duration and
the accumulated time) will be
automatically updated.
• Time: total method duration
until the end of the current step.
If the value is changed, then the
step duration (and thus the step
duration and volume and the
accumulated volume) will be
automatically updated.