6 appendix – JUMO 902930 Wtrans Transmitter RTD Temperature Probe with Ex approval Operating Manual User Manual
Page 40

6 Appendix
40
Environmental influences
Ambient temperature range of
the case
-30 to +85 °C (case including electronic)
If the operating temperature exceeds the max. ambient tempera-
ture then take into account the heat input of the protection fitting
into the case. See Chapter 1.7 "Heat input through protection tube
/ protection tube fitting", page 13.
Storage temperature range
-40 to +85 °C (case including electronic components)
Temperature influence
≤
±0.0025 %
1
/K; per K deviation
from the reference temperature 22 °C (±3 K) (case)
1
All accuracy specifications in % with regard to the measuring range of 290 K.
Resistance to climatic conditions Rel. humidity
≤
95 % without condensation
according to IEC 68-2-30
Vibration resistance
Max. 2 g at 10 to 2000 Hz (relating to the case including electronic
components) according to DIN IEC 60068-2-6
Admissible mechanical
shock resistance
25 g/6 ms (with regards to the case including electronic compo-
nents) DIN IEC 68-2.29 per 1000 cycles
EMC
- Interference emission
- Interference immunity
- Radio frequency spectrum
DIN EN 61326-1
Class B
Industrial requirements
ETSI EN 300 220-1, ETSI EN 300 220-2
Case
Material
PA6 (polyamide), color: blue
Flammability class
UL 94 HB
Dimensions
Diameter Ø approx. 32 mm, length approx. 126 mm;
for insertion lengths refer to order details
Protection type
IP67 according to DIN EN 60529;
for basic type 902930/15 and 902930/17;
for basic type 902930/55 only with screwed-on M12 × 1 machine
connector
Installation position
Any
Weight
Approx. 120 g (without protection fitting)
Ex identification
Zone
Identification marking
Use in zone 0 (20)
TÜV 11 ATEX 555248
II 1 G Ex ia IIB T4 Ga
II 1 D Ex ia IIIB T130 °C Da
-30 °C
≤
Ta
≤
+85 °C
Use in zone 0 (20) / zone 1 (21)
with zone separation by protec-
tion fitting
TÜV 11 ATEX 555248
II 1/2 G Ex ia IIB T4 Ga/Gb
II 1/2 D Ex ia IIIB T130 °C Da/Db
-30 °C
≤
Ta
≤
+85 °C