6 appendix, 1 technical data, Appendix – JUMO 902930 Wtrans Transmitter RTD Temperature Probe with Ex approval Operating Manual User Manual
Page 39: Technical data

39
6 Appendix
6.1
Technical data
Analog input
Measuring input
Pt1000 according to DIN EN 60751, in 3-wire circuit
Operating temperature range
-30 to +260 °C
Accuracy
of the temperature sensing ele-
ment
Class A ±(0.15 K + 0.002 × ItI)
ItI = measured temperature in °C without prefix
Sensor lead resistance
≤
11 ohm per lead with 3-wire circuit
Sensor current
≤
500 µA
Measuring circuit monitoring
Detection of probe break and probe short-circuit
Output (radio transmission)
Transmitter detection
(transmitter ID)
Max. 5-digit ID, factory-set, cannot be configured
Transmission interval
Adjustable via DIP switch to 5 s, 10 s, 20 s, or 45 s
(factory-setting = 20 s)
Radio frequency
868.4 MHz
Transmission power
< +10 dBm
Open air range
300 m and use of the receiver
antenna wall holder, and 3 m antenna cable.
When installing the antenna directly on the receiver, the user must
take approx. 40 % less range into consideration.
Output signal
882.2 to 1977.1 ohm = -30 to +260 °C
Response time
of the complete probe
In water 0.4 m/s / in air 3.0 m/s
∅
4.0 mm:
Water t
0.5
approx. 3 s, t
0.9
approx. 7 s / air t
0.5
approx. 25 s, t
0.9
approx. 80 s
∅
6.0 mm:
Water t
0.5
approx. 4 s, t
0.9
approx. 10 s / air t
0.5
approx. 32 s, t
0.9
approx. 98 s
Calibration accuracy
of the electronic components
≤
±0.05 %
1
1
All accuracy specifications in % relating to the measuring range of 290 K.
Voltage supply
Lithium battery
Voltage: 3.6 V, rated capacity: 2.2 Ah
Operating life
Approx. one year with the factory-set values and at room tempera-
ture (a fast transmission interval and a high or low ambient temper-
ature reduce the battery operating life)
Battery change
Only use the preconfigured lithium battery that is available in the
accessories (sales no. 90/00525539)