background image

tinb01e1-f (2011-07) VGC401.om

75


"Transmit (T)" and "Receive (R)" are related to the host.

T:

TID

[]

R:
T:
R: PSG

Request for gauge identification
Positive acknowledgement
Request for data transmission
Gauge identification

T:

SP1

[]


R:
T:
R: 1.0000E-09,9.0000E-07

Request for parameters of
switching function (setpoint)
Positive acknowledgement
Request for data transmission
Thresholds

T:

SP1

,6.80E-3,9.80E-3 []


R:

Modification of threshold values of
switching function (setpoint)
Positive acknowledgement

T:

FOL

,2 []


R:
T:
R: 0001
T:

FIL

,2 []

R:
T:
R: 2

Modification of filter time constant
(syntax error)
Negative acknowledgement
Request for data transmission
ERROR word
Modification of filter time constant
Positive acknowledgement
Request for data transmission
Filter time constant

T:

PR1

[]

R:
T:
R: 0,8.3400E-03
T:
R: 1,8.0000E-04

Request for measurement data
Positive acknowledgement
Request for data transmission
Status and pressure
Request for data transmission
Status and pressure