beautypg.com

INFICON VGC401 Single-Channel Controller User Manual

Page 39

background image

tinb01e1-f (2011-07) VGC401.om

39

Available for:

†

Pirani

(PSG)

†

Pirani/Capacitive

(PCG)

†

Cold cathode

(PEG)

†

Cold cathode/Pirani

(MPG)

†

Hot cathode

(BAG)

†

Hot cathode/Pirani

(BPG, HPG)

;

Capacitive

(CDG)

†

Hot cathode/Pirani/Capacitive

(BCG)


When the offset correction is activated, the stored offset
value is subtracted from the actual measurement value.
This allows measuring relative to a reference pressure.

The offset correction affects:

;

the displayed measurement value

†

the displayed threshold value of the switching

functions

†

the analog output at the CONTROL connector

(

23)


Value

Ö

Offset correction

deactivated

e.g.:

Ö

Offset correction

activated

para

Ö

Press briefly:

Activate offset correction.

para

Ö

Press key >2 s:

The offset value is read-
justed. The actual measure-
ment value is accepted as
new offset value.

para

Ö

Press briefly: