68

tinb01e1-f (2011-07) VGC401.om

Transmit:

COR

[,[x]x.xxx] []


└─ 0.100 … 10.000

(default = 1.000)

Receive:
Transmit:
Receive: [x]x.xxx


└─ Correction factor


Transmit:

DCD

[,x] []


└─ x = 2 –> 2 digits (default)

3 –> 3 digits

Receive:
Transmit:
Receive: x


└─

Number of digits


When the PrE (

→ 46) is ON and the pressure is in the

range p<1.0E-4 mbar the display resolution of the PSG
and PCG Gauges is reduced by one decimal digit.


Transmit:

FIL

[,x] []


└─ x = 0 –> fast

1 –> medium (default)

2 –> slow

Receive:
Transmit:
Receive: x


└─ Filter time constant


Correction factor

Number of digits in
the display

Measurement value
filter