APC driver installation

All standard APC drivers are installed in the detector.

Note:

Battery operated model Sensistor ISH2000 does not incorporate the APC
feature.

For custom APC driver design and installation contact support at INFICON.

8.17 Sensistor ISH2000P installation

The panel mount model can be installed in the operator’s panel or any other flat
surface of your leak tester. Mounting brackets and panel rubber seal are delivered
with the detector. Refer to Figure 8-6.

Fig 8-7. Panel cut out dimensions.

Measurements:

Panel cut out: 262 x 127 mm (5 x 10.3 inches).

Maximum panel thickness: 8 mm (0.3 inch).

Allow an extra 20 mm (0.8 inch) on left and right side for mounting brackets.

Depth of open space in panel is at least 15 cm (6 inches) for Sensistor ISH2000 to fit.

CAUTION!

The detector should preferably be mounted on a
vertical surface. Tilting more than 30 degrees is not
recommended. Tilting more makes air circulation
poor resulting in increased temperature inside
detector. This will reduce contrast of display and
lifetime of lamp and electronic circuits.

NOTICE!

After installation verify that ambient
temperature is below 50ºC.

Free space horisontally
300 mm (12 inches)

262 mm (10.3 inches)

127 mm (5 inches)

Free space vertically
140 mm (5.5 inches)