Min Presentation Time

Signal values in

Measuring Mode

will never be presented shorter than this time.

Values are, however, always presented until the signal has recovered. The default
value is 1 second, but values from 0 - 120 seconds can be used.

Display Threshold

Hides all measurements under a set % of

Reject Level

.

Audio Threshold

Instrument is silent under a set % of

Reject Level

.

Reject Indications

There are three choices of

Reject Level

indications except the LEDs indication:

Flashing screen

Chopped audio signal

Combination of indication 1 and 2.

Show Reject Level

Shows the Reject Level value on the display.

Audio Ready Pulse

This sets the standby sound to a silent or pulsating tone.

8.9 APC Settings*

APC is an abbreviation for Active Probe Control. APC is used to control an active
probe via the Probe Control Port. There are several active probes available. Choose
correct probe under the menu Probe Type. Customized APC drivers can be
downloaded from a PC.

There is a possibility to adapt how to measure by adjusting the timers and

Purge

Level

.

* Not in Battery Mode

Probe Type

Select the connected probe. When a custom driver is downloaded, it will replace
No driver installed”.

APC Time A-D

Adjustable timer used by the APC system. APC timer can be used for general
purposes in a custom APC program.

Purge Level

Standard probes that support active sampling use the Purge Level for fast
interruption of sampling.

Setting

Purge Level

equal to, or just above,

Reject Level

will give the fastest possible

cycle times for those probes.