Hydrogen leak detector Sensistor ISH2000
-
Operating Manual
OM-Sensistor ISH2000-EN-201304
33
EN
Measure/Print Button
Setting this parameter to
ON
displays
Measure
or
above the button 1.
Measure
will be displayed for an APC-Probe or
for a Hand Probe. Pressing
Measure
will
initiate a sample cycle. Pressing
will send the values from the hand probe
measurement to the printer port.
Probe Button
This is for setting the different functions with the probe button. These functions are
as follows:
•
Toggle Mode-makes it possible to switch between Measuring mode and
Locating mode.
•
Zero Locating signal in Locating mode.
•
Measure/Print- makes it possible to initiate sample cycles or send the values
from the hand probe measurement to the printer port.
•
Probe Lamp- makes it possible to turn on and off the Probe Lamp.
Probe Lamp
Makes it possible to have the Probe Lamp ON even if the other Probe Button
function is chosen.
Change Password
The user password is a text string (max 12 alphanumerical characters) used to lock
critical parameters. Setting password to an empty string (no characters) means that
no password is needed to modify the critical parameters. The default is no password
(“”).
Contact INFICON AB if you have lost your user password. If the
Password Protected
Calibration
parameter is set to
ON
you will be prompted for a password when starting
a calibration.
Note:
Setting
Password Protected Calibration
to
ON
has no effect if no password is set.
Note:
APC controlled calibration can be started from the bus in both cases.
Password Protection Level
Password Protection Level sets the protection level.
Standard is used when the main parameters shall be protected. High is used to
protect all settings, such as sound and display properties.
Description of the different Password Protection Level
Password
protected
settings/
functions
Non
(No password has
been set)
Standard
Standard
+
Password
Protected
Calibration
High
High
+
Password
Protected
Calibration
Change Test Mode
Yes
Yes
Calibration
(function)
Yes
Yes