Hydrogen leak detector Sensistor ISH2000

-

Operating Manual

OM-Sensistor ISH2000-EN-201304

33

EN

Measure/Print Button

Setting this parameter to

ON

displays

Measure

or

Print

above the button 1.

Measure

will be displayed for an APC-Probe or

Print

for a Hand Probe. Pressing

Measure

will

initiate a sample cycle. Pressing

Print

will send the values from the hand probe

measurement to the printer port.

Probe Button

This is for setting the different functions with the probe button. These functions are
as follows:

Toggle Mode-makes it possible to switch between Measuring mode and
Locating mode.

Zero Locating signal in Locating mode.

Measure/Print- makes it possible to initiate sample cycles or send the values
from the hand probe measurement to the printer port.

Probe Lamp- makes it possible to turn on and off the Probe Lamp.

Probe Lamp

Makes it possible to have the Probe Lamp ON even if the other Probe Button
function is chosen.

Change Password

The user password is a text string (max 12 alphanumerical characters) used to lock
critical parameters. Setting password to an empty string (no characters) means that
no password is needed to modify the critical parameters. The default is no password
(“”).

Contact INFICON AB if you have lost your user password. If the

Password Protected

Calibration

parameter is set to

ON

you will be prompted for a password when starting

a calibration.

Note:

Setting

Password Protected Calibration

to

ON

has no effect if no password is set.

Note:

APC controlled calibration can be started from the bus in both cases.

Password Protection Level

Password Protection Level sets the protection level.

Standard is used when the main parameters shall be protected. High is used to
protect all settings, such as sound and display properties.

Description of the different Password Protection Level

Password
protected
settings/
functions

Non
(No password has
been set)

Standard

Standard
+
Password
Protected
Calibration

High

High
+
Password
Protected
Calibration

Change Test Mode

Yes

Yes

Calibration
(function)

Yes

Yes