3 commands, Commands – INFICON HLD6000 Refrigerant Leak Detector Interface Description User Manual
Page 8

8
ASCII Protocol
2.3
Commands
Command
Meaning
Relates
to LD
cmd. no.
Read /
Write
*BEEP
Output beep sound. Parameter: Typ,Volume
"*BEEP 1,2"
Typ: 1= 2000 Hz, 2=Dual (2000 and 1000 Hz), 3= 1000 Hz
Volume: Volume between 1 and 15
W
*CAL
Start external calibration
4
W
*CAL:ACKnowledge
Acknowledge stable signal during external calibration
11
W
*CLS
Clear Error
5
W
*CONFig:AUDio
Audio alarm type (PIN, SET, TRIG, OFF)
600
R/W
*CONFig:BRIGHTness
Sniffer white LED brightness
414
R/W
*CONFig:BUTSniffer
button of the sniffer probe (OFF, ON)
422
R/W
*CONFig:CALleak
leak rate of external test leak
R/W
*CONFig:CALleak:G/a
leak rate of external test leak in g/a
388
R/W
*CONFig:CALleak:LB/yr
leak rate of external test leak in lb/yr
R/W
*CONFig:CALleak:MBAR*l/s
leak rate of external test leak in mbar*l/s
R/W
*CONFig:CALleak:OZ/yr
leak rate of external test leak in oz/yr
R/W
*CONFig:CALleak:PA*m3/s
leak rate of external test leak in Pa*m3/s
R/W
*CONFig:CALREQInterval
calibration request interval in minutes (0=calibration
request off)
418
R/W
*CONFig:GAS
Gas type (smart probe only)
(R134A, R22, R404A, R407C, R410A, R1234YF, R32,
USER1, USER2, USER3)
486
R/W
*CONFig:LIGHTAlarm
Sniffer white LED alarm configuration:
DISABLED: always same brightness
ON: higher brightness if trigger 1 is exceeded
BLINK: blink, if trigger 1 is exceeded
413
R/W
*CONFig:PLCINLINK
Assignment of PLC-inputs
*CONFig:PLCINLINK:1 for PLC input Pin1
*CONFig:PLCINLINK:2 for PLC input Pin2
...
*CONFig:PLCINLINK:10 for PLC input Pin10
438
R/W
*CONFig:PLCINLINK:1
Configuration of PLC-input. The following settings are
possible: NOT_USED, EXT_CAL, START, STOP,
CLEAR, START_STOP, VALID_TRIGGER.
438
R/W
*CONFig:PLCINLINK:n
n = 2…10 see CONFig:PLCINLINK:1
438
R/W
*CONFig:PLCOUTLINK
Assignment of PLC-outputs
*CONFig:PLCOUTLINK:1 for Pin 1
*CONFig:PLCOUTLINK:2 for Pin 2
...
*CONFig:PLCOUTLINK:8 for Pin 8
263
R/W
*CONFig:RECorder:LINK1
Function at analog output channel 1
(OFF, LR_LIN, EXTERN)
222
R/W
*CONFig:RECorder:LINK2
Function at analog output channel 2
(OFF, LR_LIN, EXTERN)
222
R/W
*CONFig:RECorder:UPPERVAL
Upper limit for the analog out at IO module.
The value sets the leak rate in g/a for an output voltage of
10V
225
R/W
*CONFig:RS232
Protocol (ASCII, LD, HLD4000, SIMPLE)
26
R/W