4 acyclic data transfer, 1 addressing rules for acyclic access, Acyclic data transfer 4.4.1 – INFICON HLD6000 Refrigerant Leak Detector Interface Description User Manual
Page 35: Addressing rules for acyclic access

Fieldbus Communication
35
4.4
Acyclic Data Transfer
If you want to use acyclic data transfer with PROFIBUS, you must use a PROFIBUS
master which supports DPV1 data transfers.
A PROFIBUS master which supports DPV0 only, can only use cyclic data transfer.
4.4.1
Addressing Rules for Acyclic Access
Mapping from LD command number to field bus
Fieldbus supports all commands from LD protocoll, except the commands in the
following list:
Fieldbus
Rule
Example for LD_command_number 506
(Mass)
PROFIBUS
LD_command_number = slot * 255 + index +1
slot = (ADI-1) / 255
index = (ADI-1) MOD 255
Slot = 1
index = 250
PROFINET IO Application Process Instance (API) = 0
Slot = 0
Subslot = 1
Index = LD_command_number
API = 0
Slot = 0
Subslot = 1
Index = 506dez = 01FAhex
DeviceNet
Object number A2h (ADI object)
Instance_number = LD_command_number
Instance_number = 506
EtherNet/IP
Object number A2h (ADI object)
Instance_number = LD_command_number
Instance_number = 506
LD_Command number
Name
26
Interface protocol (read only)
27
Used interface
275
Cal history
287
Error history
294
Text of error number
295
Text of warning bit
341
Last entry in long term history list
342
Long term history list
408
Serial number IO module
423
Speaker beep
426
Front panel LED state
450
Date+Time [YMDhms]
457
COOL-Check information
458
Probe HEX code
619
Pump power attenuation
1145
Gas correction factor
1161
Parameter reset