INFICON LDS3000 Modular Leak Detector Interface description User Manual
Page 45
LD Protocol
45
Cal history (command 275)
Error history (command 287)
TMP error history (command 288)
Present warnings (command 297)
Answer
ListNo, 'Fac:', Calfac(float), 'Leak:', Testleak(float),
'Anod:', Anodevoltage, 'M', Mass, 'VAC' or 'SNIF',
year/month/day, hour:min:sec,
'Cat:', Cathode, 'State:', cal state
Example
08 Fac: 0.00E+0 Leak: 0.00E+0
Anod: 000 M2 VAC
2000/00/00 00:00:00
Cat: 1 State: 000
Answer
ListNo, 'ERR' or 'WRN', ErrNo, ErrValue(float),
year/month/day, hour:min:sec,
'SwOnCnt:', SwitchOnCnt, 'OnTm:', MinSinceStart
Example
05 WRN220 2.103E+1
2012/03/26 09:27:48
SwOnCnt: 028 OnTm: 015
Answer
ListNo, 'ERR' or 'WRN', ErrNo
Example
05 WRN220
Value
Meaning
0x00000001
Warning pressure/flow
0x00000002
Warning pressure rise
0x00000004
Warning anode voltage
0x00000008
Warning pirani
0x00000010
Warning emission
0x00000020
Warning suppressor
0x00000040
Warning TMP
0x00000080
Warning Anybus
0x00000100
Warning maintenance
0x00000200
Warning I/O disconnected
0x00000400
Warning 5V
0x00000800
Warning U24VHz
0x00001000
Warning U24V Pwr12
0x00002000
Warning U24V Pwr 34
0x00004000
Warning U24V Pwr 56
0x00008000
Warning U24V8
0x00010000
Warning U24V9
0x00020000
Warning U24V10
0x00040000
Warning U24V11
0x00080000
Warning cathode voltage
0x00100000
Warning temperature MSB
0x00200000
Warning temperature preamplifier
0x00400000
Warning calibration request
0x00800000
Warning sniffer not connected
0x01000000
Preamp output too low