LD Protocol

39

432

1B0

Leak rate unit
sniff

R/W UINT8

0, 0, 5

Leak rate unit sniff
0 - mbarl/s
1 - Pam3/s
2 - Atm ccs
3 - Torrl/s
4 - ppm
5 - g/a

433

1B1

Anode setpoint
M2 [V]

R/W UINT16

785, 905, 995

Anode voltage setpoint for mass 2
(hydrogen) in V

434

1B2

Anode setpoint
M3 [V]

R/W UINT16

510, 610, 670

Anode voltage setpoint for mass 3 in V

435

1B3

Anode setpoint
M4 [V]

R/W UINT16

390, 465, 520

Anode voltage setpoint for mass 4
(helium) in V

436

1B4

Emission
current setpoint
[V]

R/W FLOAT

1E-4, 2.5E-3,
2.8E-3

Emission current setpoint [V]

438

1B6

PLC input
configuration
IO module

R/W UINT8[10]

PLC_IN 1: -16, 11,
16
PLC_IN 2: -16, 4,
16
PLC_IN 3: -16, -12,
16
PLC_IN 4: -16, 7,
16
PLC_IN 5: -16, 2,
16
PLC_IN 6: -16, 3,
16
PLC_IN 7: -16,
9,16
PLC_IN 8: -16, 0,
16
PLC_IN 9: -16, 0,
16
PLC_IN 10: -16, 0,
16

Configuration of PLC input port of the IO
module
Index 0...9 = PLC_IN1...PLC_IN10
See table "PLC input conf."

439

1B7

Key switch
state

R

UINT8

Key switch state
0=inactive, 1=active, 2= not used
Bit0&1: KEY_1
Bit2&3: KEY_2
Bit4&5: KEY_3
Bit6&7: not used

448

1C0

Valve control
location

R/W UINT16

Bit=0: Controled by leak detector
Bit=1: Controled by write command 449

449

1C1

Switch valves

R/W UINT16

see table "Valves"
For setting valve by write comman see
also command 448

Comm
and dez

Com
mand
hex

Name

R/W

Data type

Min-, Def.-, Max-
value

Meaning