Disassemble, forward & store, Disposal, Troubleshooting – Heidolph VAC control automatic User Manual
Page 65

65
E
DISASSEMBLE, FORWARD & STORE
Disassemble
Turn item OFF and disconnect mains plug.
Transportation and store
1. Store item and its components in the original box to avoid damage in transit. Close box with
adhesive tape.
2. Store item in a dry place.
Caution: avoid shocks an vibration in transit.
DISPOSAL
For disposal, please comply with your local or national regulations.
Split by metal, plastic, etc.
Packing material to be treated as described above (material split).
TROUBLESHOOTING
Vacuum pumps
defect found
reason
correction
pump won't run or
starts running and
stops right away
electricity missing, voltage too
low ?
overpressure in exhaust line?
motor overload?
check voltage, connect
electricity
open exhaust line.
allow motor to chill, check for
reason, turn ON again
low suction power
centering ring not seated
correctly
long, thin line?
condensate in pump?
deposits in pump ?
diaphragm and valves broken?
emission of gas, vapor in
process ?
check pump check fittings. line
and recipient
increase cross section.
run pump with intake stud
open for a minute
clean and check pump heads
replace diaphragm and/or
valves
check process parameters
pump noisy
sucks at atmospheric
pressure ?
cocking disc loose?
none of the above reasons
found ?
install hose on exhaust end
maintain diaphragm pump
ship pump for repair.
pump stuck or piston
rod jammed
-
ship pump for repair