down onto the Teflon ferrule. This clamps the Teflon ferrule onto the reagent line
and, if done properly, insures a pressure tight connection.

2.2.3 Reagents

The Reagent Inlet Ports of your AutoBlock have been initially set up in the
following manner:

Port #1: DI Water

DI Water must always be attached to Port #1.

Port #2: Hydrochloric Acid
Port #3: Nitric Acid
Port #4: Hydrogen Peroxide
Port #5: Sulfuric Acid
Port #6-10 for dual system only-

DI Water must always be attached to Port #6

for dual sysems

If these reagents do not suit your application, you may change them. Changing
the reagents and port numbers is addressed in Chapter 6, Service Mode, Setup
Section.


2.3

Install the AutoBlock
This is an easy process but be careful to follow these steps to insure proper
function of the unit. The rear panel of your AutoBlock should continue to remain
accessible during this section.

2.3.1 Power Cord-The AutoBlock includes a 6’ power cord. Power requirements are

standard 115VAC 12amp US and Canada, 220VAC 6amp International. Plug the
power cord into the rear panel port labeled POWER IN, and plug into a standard
outlet. If the power line must be extended beyond 6’, the use of an extension cord
with the appropriate minimum electrical and safety standards may be used.


2.3.2 Remove the PDA from the box. Place it in the AutoBlock PDA cradle facing

forward, aligning the gold plated pins. The PDA cradle is mounted permanently
on the right side of the AutoBlock.


2.3.3 Attach the cord originating in the PDA cradle to the serial port labeled RS 232 on

the rear panel.


2.3.4 The PDA Power Cable is located in the HP Jornada PDA box. It has a 2-prong

plug and a black AC/DC adapter box. Plug this into the rear panel of the
AutoBlock in the receptacle labeled PDA.


2.3.5 There is a second cord in the PDA box. It is 4.5’ in length. One end of this cord

has 2 gold tabs; the other end is an “L” shaped plug. Attach this cord into the
pigtail coming from the RS 232 cable, and attach the “L” shaped plug into the
AC/DC adapter box. (Note: some of these cables may already be assembled by
the PDA manufacturer) When the PDA is place on the cradle a light should
illuminate on the top of the PDA. Red means the PDA is charging, Green is the
batteries are fully charged. If no light appears check all connections making sure