4.4. Wiring & Power Connections
See electrical schematics in Section 8.3 for line connections.
The oven must be hardwired directly to the electric supply.
Verify proper fan rotation on initial start-up and recheck rotation
whenever wiring or electrical component changes are made.
Each fan has an arrow showing correct rotation.
Danger!
All grounding and safety equipment must be in compliance with
applicable codes, ordinances and accepted safe practices.
Notice
Flashing through roof or wall must be capable of handling
temperatures up to 343°C (650°F).
All stacks must comply with state and local building codes to
ensure that surrounding combustible surfaces are below 71°C
(160°F).