2 auto probe recognition, Probe automatically recognized – Dakota Ultrasonics CMX DL plus User Manual
Page 9
![background image](/manuals/557255/9/background.png)
CMX
DL+
High Performance Material & Coating Thickness Gauge
5
2.2 Auto Probe Recognition
When the CMX
DL+
is initially powered up, the gauge will automatically check to see if
the transducer plugged into the gauge can be recognized. The steps that follow
assume the CMX
DL+
recognized the probe type:
Probe Automatically Recognized
1) Press the OK key once to use the identified probe, or ESC to display a list of
optional transducers. Note: if the
CMX
DL+
recognizes a specific
transducer, the user should always select OK to use the identified probe.
The only time an alternative probe should be selected from a list is if the
user switched probes following initial power up and recognition.
2) Assuming the
CMX
DL++
recognized the probe and the OK key was pressed,
the
CMX
DL+
will advance to a Zero Probe menu. If the transducer was
identified as a special transducer capable of measuring coating thickness, a
menu will be displayed allowing the user the ability to toggle the coating
thickness display on/off as follows: