Dakota Ultrasonics CMX DL plus User Manual

Page 27

background image

CMX

DL+

High Performance Material & Coating Thickness Gauge

23

RF View

Rectified (RECT) View

BSCAN View

DIGITS View

In the upper left corner of each of the display photos above, is the repeatability
indicator. The repeatability indicator is represented by six vertical bars and
represents how repeatable the measurements are. In regular measurement mode,
the CMX

DL+

makes 8 measurements a second. In scan mode, the CMX

DL+

makes

200 measurements a second. If the coating mode option is activated, the CMX

DL+

makes 3 measurements a second in regular measurement mode and 65
measurements a second in scan mode. When the CMX

DL+

is idle, only the left

vertical bar will be displayed. However, when the CMX

DL+

is making a measurement,

four or five of the bars should be displayed on the repeatability indicator. If fewer
than four bars are showing, the CMX

DL+

is having difficulty achieving a stable

measurement and the thickness value displayed is potentially unstable.