beautypg.com

Dakota Ultrasonics CMX DL plus User Manual

Page 110

background image

Dakota Ultrasonics

106

3) Use the UP and DOWN arrow keys to scroll through the sub menu items

until COATING 1PT is highlighted.

4) Press the ENTER key to display the Digits Edit Box.

5) Press the UP and DOWN arrow keys to scroll the highlighted value.

6) Press the LEFT and RIGHT arrow keys to scroll the digit locations.

7) Repeat steps 5 & 6 until the known thickness value is correctly displayed.

8) Press the OK key to calculate the velocity and return to the menu screen, or

ESC to cancel the coating one point calibration.

9) Finally, press the MEAS key to return to the measurement screen and begin

taking readings.