Adjusting the range (b-depth), Adjusting the range (b-depth) using the hot menus – Dakota Ultrasonics CMX DL plus User Manual

Page 75

background image

CMX

DL+

High Performance Material & Coating Thickness Gauge

71

5) Press the LEFT and RIGHT arrow keys to scroll the digit locations.

6) Repeat steps 4 & 5 until the DELAY value is correctly displayed.

7) Press the OK key to return to the measurement screen, or ESC to cancel

entering the DELAY.

8) Finally, press the MEAS key to return to the measurement screen and begin

taking readings.

Adjusting the Range (B-Depth)

Once again, just to reiterate, the Range (B-Depth) + Delay (B-Start) equals the
right side of the screen
. Therefore, the Range (B-Depth) is the overall area, from
the delay, that will be viewable on the screen. The CMX

DL+

digitizer will round up

from the Range (B-Depth) that’s entered. Therefore, if the Range (B-Depth) is set at
1.0”, the digitizer will round this value up to the next adjustment available. The
procedures to adjust the overall thickness range viewed Range (B-Depth) are
outlined below:

Adjusting the Range (B-Depth) using the Hot Menus

1) Press the MEAS key once to activate measure menu items. Press the

MEAS key multiple times to move right and the ESC key multiple times to

move left, until the WIDTH cell is highlighted.