Eppendorf 5430 R Centrifuge User Manual
Page 56
Troubleshooting
Centrifuge 5430/5430 R
English (EN)
56
Error 16
to
Error 17
Electronics fault
Electronics fault.
Switch centrifuge off and back on
again after > 20 s.
Error 18
Room Temp. of
rotor chamber
(only 5430 R)
Temperature deviation from set value in
the rotor chamber: ΔT > 16 °C.
Allow the device to cool down and
repeat cycle.
Error 18
Room Temp. of
rotor chamber
(only 5430 R)
Temperature deviation from set value in
the rotor chamber: ΔT > 50 °C.
Allow the device to cool down and
repeat cycle.
Error 22
Electronics fault
(only 5430 R)
Electronics fault.
Switch centrifuge off and back on
again after > 20 s.
Error 25
Mains/power failure
Mains/power failure during a run.
Check the power supply.
Error 26
Electronics fault
(only 5430 R)
Electronics fault.
Switch centrifuge off and back on
again after > 20 s.
Error 27
Electronics fault
(only 5430 R)
Electronics fault.
Switch centrifuge off and back on
again after > 20 s.
Error 28
Electronics fault
Electronics fault.
Press the open key.
Error 30
Lid latch
Centrifuge lid could not be locked.
Try again to close centrifuge lid.
Error 30
Lid latch
Centrifuge lid could not be released.
Switch the device off and back on
again.
If the error occurs again:
1.
Switch off the device.
2.
Activate the emergency lid release
(see Emergency release on p. 57).
Error 30
Lid latch
Centrifuge lid has not been opened wide
enough.
Open the centrifuge lid wider by
hand.
Symptom/message Cause
Remedy