beautypg.com

Eppendorf 5430 R Centrifuge User Manual

Page 56

background image

Troubleshooting
Centrifuge 5430/5430 R
English (EN)

56

Error 16

to

Error 17

Electronics fault

Electronics fault.

 Switch centrifuge off and back on

again after > 20 s.

Error 18
Room Temp. of
rotor chamber

(only 5430 R)

Temperature deviation from set value in
the rotor chamber: ΔT > 16 °C.

 Allow the device to cool down and

repeat cycle.

Error 18
Room Temp. of
rotor chamber

(only 5430 R)

Temperature deviation from set value in
the rotor chamber: ΔT > 50 °C.

 Allow the device to cool down and

repeat cycle.

Error 22
Electronics fault

(only 5430 R)

Electronics fault.

 Switch centrifuge off and back on

again after > 20 s.

Error 25
Mains/power failure

Mains/power failure during a run.

 Check the power supply.

Error 26
Electronics fault

(only 5430 R)

Electronics fault.

 Switch centrifuge off and back on

again after > 20 s.

Error 27
Electronics fault

(only 5430 R)

Electronics fault.

 Switch centrifuge off and back on

again after > 20 s.

Error 28
Electronics fault

Electronics fault.

 Press the open key.

Error 30
Lid latch

Centrifuge lid could not be locked.

 Try again to close centrifuge lid.

Error 30
Lid latch

Centrifuge lid could not be released.

 Switch the device off and back on

again.

If the error occurs again:
1.

Switch off the device.

2.

Activate the emergency lid release
(see Emergency release on p. 57).

Error 30
Lid latch

Centrifuge lid has not been opened wide
enough.

 Open the centrifuge lid wider by

hand.

Symptom/message Cause

Remedy

This manual is related to the following products: