beautypg.com

4 rotors, Rotors – Eppendorf 5430 R Centrifuge User Manual

Page 12

background image

Product description
Centrifuge 5430/5430 R
English (EN)

12

2.4

Rotors

The Centrifuge 5430/5430 R can be operated with the following rotors. Before using tubes, observe the
manufacturer's recommended specifications on resistance to centrifugation (max. g-force).

Max. capacity

Max. g-force
(rcf)/speed
(rpm) without
adapter

Max.
load per
rotor
bore

(1)

Notes

Acceleration/
deceleration time

(2)

(soft): with soft ramp

Rotor
FA-45-48-11
With aerosol-tight
QuickLock rotor lid

48 tubes 1.5/2.0 mL.
With adapters:

• 0.2 mL PCR tubes
• 0.4 mL tubes
• 0.5 mL tubes
• 0.6 mL Microtainer

Outer ring:
18,213 × g
Inner ring:
16,048 × g/
12,700 rpm

3.75 g

• Aerosol-tight

(3)

QuickLock rotor lid
(aluminum).

≤ 20 s/20 s
≤ 61 s/65 s (soft)

Rotor
F-45-48-11
With polypropylene
rotor lid

48 tubes 1.5/2.0 mL.
With adapters:

• 0.2 mL PCR tubes
• 0.4 mL tubes
• 0.5 mL tubes
• 0.6 mL Microtainer

Outer ring:
18,213 × g
Inner ring:
16,048 × g/
12,700 rpm

3.75 g

≤ 20 s/20 s
≤ 61 s/65 s (soft)

Rotor
FA-45-30-11
With aerosol-tight
QuickLock rotor lid

30 tubes 1.5/2.0 mL.
With adapters:

• 0.2 mL PCR tubes
• 0.4 mL tubes
• 0.5 mL tubes
• 0.6 mL Microtainer

20,817 × g/
14,000 rpm

3.75 g

• Aerosol-tight

(3)

QuickLock rotor lid
(aluminum).

• PTFE-coated

(particularly resistant
to chemicals),
marked: coated

• Spin columns

available, better with
rotor FA-45-24-11-kit.

≤ 15 s/15 s
≤ 61 s/65 s (soft)

Rotor
F-45-30-11
With polypropylene
rotor lid

30 tubes 1.5/2.0 mL.
With adapters:

• 0.2 mL PCR tubes
• 0.4 mL tubes
• 0.5 mL tubes
• 0.6 mL Microtainer

20,817 × g/
14,000 rpm

3.75 g

≤ 15 s/15 s
≤ 61 s/65 s (soft)

This manual is related to the following products: