At 24 11 – Eppendorf 5430 R Centrifuge User Manual
Page 15
Product description
Centrifuge 5430/5430 R
English (EN)
15
(1) Maximum load per rotor bore for adapter + tube + contents.
(2) According to DIN 58 970 (device versions: 230 V, 120 V and 100 V, 50 to 60 Hz).
(3) Aerosol tightness tested and certified by the Centre of Emergency Preparedness and Response, Health
Protection Agency, Porton Down (UK) (see certificates at the end of this operating manual).
For the rotors and rotor lids labeled coated, color fluctuations may occur as a result of the production
process. These fluctuations have no effect on service life or resistance to chemicals.
Rotor
F-35-6-30
6 conical tubes 50 mL, with or
without skirted bottom
or
6 conical tubes 15 mL, with
enclosed adapters
or
6 Centriplus centrifuge filter
units with adapters.
7,745 × g/
7,830 rpm
110 g
• The rotor can only be
removed and inserted
using the supplied
removal tool.
• Centrifugation of
round-bottom tubes
and blood taking
systems can be
completed using
additional adapters
(see p. 86).
≤ 23 s/23 s
≤ 62 s/67 s (soft)
Rotor
A-2-MTP
Two buckets to hold:
• Microplates
• Cell culture plates
• PCR plates
• Deepwell plates (max.
height 29 mm)
• Slide (with CombiSlide
adapter)
2,204 × g/
4,680 rpm
170 g
(per
bucket)
• PCR plates can only
be centrifuged with
appropriate adapters.
• Max. loading height:
29 mm.
•
Only 5430 R: More
efficient cooling via
centrifugation without
wind shield upper
shell (see
Swing-bucket rotor on
p. 34).
≤ 18 s/21 s
≤ 63 s/67 s (soft)
Max. capacity
Max. g-force
(rcf)/speed
(rpm) without
adapter
Max.
load per
rotor
bore
(1)
Notes
Acceleration/
deceleration time
(2)
(soft): with soft ramp
Fixed-angle
rotor
Aerosol-tight
version
Angle of
bores
Number of
bores
∅ of
bores (mm)
AT
24
11
Aerosol-tight
version
Number of
bores
∅ of
bores (mm)
Swing-bucket