Eppendorf 5430 R Centrifuge User Manual
Page 17
Product description
Centrifuge 5430/5430 R
English (EN)
17
*) Centrifugation of round-bottom tubes and blood taking systems can be done using additional adapters
(see p. 86).
To determine the g-force (rcf) for a special adapter, you can perform a calculation according to DIN 58 970
with the following formula:
rcf = 1.118 · 10
-5
· n
2
· r
max
n: speed in min
-1
(rpm)
r
max
: max. centrifuging radius in cm
Example:
In Rotor FA-45-30-11, the 0.5 mL adapter has a maximum radius of 8.4 cm. At 7,000 rpm, a maximum
g-force of 4,600 × g is achieved.
Rotor F-35-6-30*
For 15 mL conical tubes
11.0
7,540
For 50 mL conical tubes
10.5
7,197
for Centriplus centrifuge filter units
11.1
7,567
A-2-MTP rotor
Without adapter
9.0
2,204
for 384 well PCR plates
7.7
1,885
for 96 well PCR plates
7.3
1,788
CombiSlide adapter
7.7
1,885
Rotor
Adapter
Max.centrifugation
radius r
max
[cm]
Max. g-force (rcf)