2.4.1
rcf display and calculation
Use the
rpm/rcf key to switch the speed of centrifugation display from rpm to g-force
(rcf). Ensure that the g-force displayed during switching is standardized to suit the
rotor in question without an adapter. When adapters are used, the following maximum
g-forces (rcf) can be achieved at maximum speed:
Rotor
Adapter
Max.centrifugation
radius r
max
[cm]
Max. g-force (rcf)
Rotor FA-45-48-11/
Rotor F-45-48-11
Without adapter
Outer ring: 10.1
Inner ring: 8.9
Outer ring: 18,210
Inner ring: 16,048
For 0.2 mL PCR tubes
Outer ring: 8
Inner ring: 6.8
Outer ring: 14,425
Inner ring: 12,261
For 0.4 mL tubes
Outer ring: 10.1
Inner ring: 8.9
Outer ring: 18,210
Inner ring: 16,048
For 0.5 mL tubes
Outer ring: 9
Inner ring: 7.8
Outer ring: 16,229
Inner ring: 14,065
For 0.6 mL Microtainer
Outer ring: 10.1
Inner ring: 8.9
Outer ring: 18,210
Inner ring: 16,048
Rotor FA-45-30-11/
Rotor F-45-30-11
Without adapter
9.5
20,871
For 0.2 mL PCR tubes
7.4
16.215
For 0.4 mL tubes
9.5
20,871
For 0.5 mL tubes
8.4
18,407
For 0.6 mL Microtainer
9.5
20,817
Rotor
FA-45-24-11-Kit
Without adapter
9.8
19,090
For 0.2 mL PCR tubes
7.7
15,000
For 0.4 mL tubes
9.8
19,090
For 0.5 mL tubes
8.7
16,950
For 0.6 mL Microtainer
9.8
19,090
Rotor FA-45-16-17
For 5.0 mL micro test tubes
9.4
21,191
Rotor S-24-11-AT
Without adapter
8.9
16,049
Rotor
FA-45-24-11-HS
Without adapter
8.8
30,130
For 0.2 mL PCR tubes
6.7
22,940
For 0.4 mL tubes
8.8
30,130
For 0.5 mL tubes
7.7
26,364
For 0.6 mL Microtainer
8.8
30,130
Rotor
F-45-64-5-PCR
For PCR strips, inside
7.7
11,987
For PCR strips, outside
8.7
13,543
Rotor
F-45-18-17-Cryo
Without adapter
9.4
8,320
For Cryo tubes
9.0
7,970