5 cooling (only 5430 r), 1 temperature adjustment, 2 temperature display – Eppendorf 5430 R Centrifuge User Manual
Page 36: 3 temperature monitoring, 4 fasttemp, Cooling (only 5430 r) 5.5.1, Temperature adjustment, Temperature display, Temperature monitoring, Fasttemp
Operation
Centrifuge 5430/5430 R
English (EN)
36
5.5
Cooling (only 5430 R)
5.5.1
Temperature adjustment
Set the temperature using the temp arrow keys between -11°C and +40°C.
You can also modify the temperature during centrifugation.
At higher ambient temperature a brief fan noise is possible until the desired temperature has been
reached. This indicates a heavy cooling performance.
5.5.2
Temperature display
The set temperature appears in the extended display.
5.5.3
Temperature monitoring
After the target temperature has been reached, the centrifuge reacts to temperature deviations during
centrifugation as follows:
5.5.4
FastTemp
This function can be used to start a temperature control run directly without samples with a rotor and
temperature-specific speed in order to quickly adjust the rotor chamber, including the rotor, buckets and
adapters, to the previously set nominal temperature.
The
FastTemp pro function for programming the temperature control run with defined start times is
described in the next section.
Prerequisites
• The centrifuge is switched on.
• The rotor and rotor lid are properly attached.
• The centrifuge lid is closed.
• Temperature and g-force (rcf)/speed (rpm) for the centrifugation are set (see Centrifuging on p. 39).
1.
Press the
fast temp key.
The display shows
Fast Temp
, the remaining duration of the temperature control run as well as the
current temperature and g-force (rcf)/speed (rpm).
The temperature control run automatically ends when the set temperature is reached. A periodic signal
tone sounds.
2.
Press the
start/stop key to terminate the temperature control run early.
After the set temperature has been reached and the temperature control run is complete, the centrifuge
keeps the rotor chamber with the centrifuge lid closed at the set target temperature if the temperature is
below the ambient temperature. However, independent of the target temperature, 4 °C must be met via
this continuous cooling in order to prevent the rotor chamber from freezing.
If the rotor is stopped:
Target temperature
During centrifugation:
Actual temperature
Deviation from set value
Action
± 3 °C
Temperatures on the display are flashing.
± 5 °C
Periodic warning tone and display
Error 18
. Centrifugation is stopped
automatically.