54

8

Troubleshooting

If you cannot remedy an error with the recommended measures, please contact your local Eppendorf
partner. The contact address can be found online at: www.eppendorf.com.

8.1

General errors

Symptom/message Cause

Remedy

No display.

No power connection.

 Check mains/power connection.

No display.

Power failure.

 Check the mains fuse of the device

(see Fuses on p. 53).

 Check the mains fuse of the

laboratory.

Lid of the device
cannot be opened.

Rotor is still running.

 Wait for rotor to stop.

Lid of the device
cannot be opened.

Power failure.

1.

Check the mains fuse of the device
(see Fuses on p. 53).

2.

Check the mains fuse of the
laboratory.

3.

Activate the emergency lid release
(see p. 57).

Device cannot be
started.

Lid of the device is not closed.

 Close device lid.

Device shakes
when it starts up.

Rotor loaded unsymmetrically.

1.

Stop the device and load
symmetrically.

2.

Restart device.

Centrifuge brakes
during a short run
centrifugation,
although the

short

key is pressed.

The

short key was released briefly more

than twice (protective function for the
drive).

 Press the short key continuously

during a short run centrifugation.

Temperature
display flashes.
(only for 5430 R)

Temperature deviation from the set
value: ±3 °C.

 Check the settings.
 Check unhindered air circulation

through the air slots.

 Thaw ice or switch off device and

allow it to cool down.