beautypg.com

7 dual wavelength, Dual wavelength – Eppendorf BioSpectrometer basic User Manual

Page 97

background image

97

Evaluation procedure

Eppendorf BioSpectrometer

®

basic

English (EN)

12.7

Dual wavelength

For methods of the

Dual Wavelength group absorbances that were measured at two wavelengths can be

calculated with each other before the calculated absorbance is evaluated further with the factor or
standard.

To determine the calculated absorbance a division or subtraction evaluation can be defined in the
parameters:

A

1

,

A

2

= measured absorbance.

a, b, c, d = factors that are entered in the parameters. Negative figures may also be entered.

d

c

A

b

A

a

A

calc

u

u

u

2

1

>

@

d

c

A

b

A

a

A

calc

u

u

u

2

1