beautypg.com

9 troubleshooting, 1 general errors, Troubleshooting 9.1 – Eppendorf BioSpectrometer basic User Manual

Page 75: General errors, 9troubleshooting 9.1 general errors

background image

75

Troubleshooting

Eppendorf BioSpectrometer

®

basic

English (EN)

9

Troubleshooting

9.1

General errors

Error

Possible cause

Remedy

Measuring results
are imprecise.

• Reagent is past its shelf life.

Ensure that the reagent is still within its shelf life
and properly prepared.

• Reagent has not been

prepared properly.

Use clean demineralized water of adequate quality
for preparation if required.

• Pipetting is not correct.

Ensure that the pipette is calibrated and that
pipetting is being performed correctly.

• Incubation procedure

before measurement is
incorrect.

If the method procedure requires incubation
before the measurement, ensure that the
temperature and time for incubation are correctly
observed.

• The cuvette is

contaminated.

Clean and rinse the cuvette. When replacing a
cuvette, pay attention that the optical window of
the cuvette remains clean and that you do not
touch it with your fingers.

If the cuvette window has become soiled from
fingerprints, wipe it clean using a lint-free lab
cloth soaked in ethanol or isopropanol.

• The cuvette is not filled

completely with measuring
solution, and it contains
bubbles.

Ensure that the required minimum volume of the
cuvette for a measurement is reached and that no
bubbles are in the measuring solution.

• Turbidity of the measuring

solution.

Centrifuge the turbid measuring solutions
containing particles and use the clear supernatant.

• Spectrophotometer is

drifting.

Contact Eppendorf Service.

Observe the ambient conditions.

Prevent temperature changes.

• Cuvette shaft is dirty.

Clean the cuvette shaft.

The measuring
results are not
correct.

• The method has not been

programmed correctly.

Ensure that the method parameters are entered
correctly.

• The standard solution has

not been prepared correctly.

Ensure that the correct standard is used and that
the measuring solution for the standard is
prepared correctly.

• The absorbance of the

reagent is drifting.

For instable reagent absorbance and end point
methods: When measuring a long series of
samples, measure the reagent blank value not only
at the beginning but also during the sample series.
If the reagent blank value drifts strongly, the
reagent is not appropriate for error-free
measurements and has to be replaced by a new
reagent.

• The cuvette is not

positioned correctly.

Position the cuvette in the cuvette shaft so that the
optical window points towards the direction of the
light path.

Photometry light path: from back to front