Troubleshooting
Eppendorf BioSpectrometer

®

basic

English (EN)

80

The coefficient of
determination for
the regression
evaluation of the
standard series is <
0.8.

• For methods with evaluation of

standard series via the regression
procedure: If the regression
evaluation for the standard series was
nonlinear, but the standard evaluation
was accepted by the user, a warning
appears after samples have been
measured.

Use the sample results with the
reservation mentioned or repeat the
measurement of the standard series
and samples.

Scan: Some of the
measured
absorbances are
too high and are
not displayed.

• For at least one scan wavelength, the

absorbance exceeded the measuring
range.

• Turbidity of the measuring solution.

• Measurements at the limits of the

photometric measuring range.

If the non-displayed areas of the scan
are relevant: Dilute the sample or
remove the dilution via centrifugation
and repeat the measurement.

Absorbance at the
measuring
wavelength is too
high.

• Turbidity of the measuring solution.

• Optical surfaces of the cuvette are

soiled.

• Cuvette has been inserted into the

cuvette shaft facing the wrong
direction.

• Too high absorbance of measuring

solution.

Measure again considering the
possible causes.

The calculated
result is negative.

• Measuring solution not prepared

correctly.

• The incorrect factor has been entered

(wrong algebraic sign).

Measure again considering the
possible causes.

At least one of the
results is negative.

• For methods with several results (e.g.,

Dye labels).

• Measuring solution not prepared

correctly.

• The incorrect factor has been entered

(wrong algebraic sign).

Measure again considering the
possible causes.

The result has more
than 6 pre-decimal
places.

• Very high sample concentration.

• Concentration unit does not match

the expected range of the sample
concentrations.

Dilute sample and measure again.

Change the concentration unit
(Parameter

Unit) and measure again.

The result is more
than 5

%

outside

of

the standard
concentration
range.

• For methods with evaluation via

standard curves (nonlinear evaluation
method):

The sample result is more than 5

%

outside of the standard concentration
range.

Remeasure the sample under
conditions under which the result is
within the range of the standard
concentrations (dilute sample, modify
standard concentrations and
remeasure).

Problem

Cause

Solution