Troubleshooting

Eppendorf BioSpectrometer

®

basic

English (EN)

• Calculation not

possible
because of
division by zero.
Absorbance
result is zero.

• Calculation

error. Division
by zero.

• The evaluation required dividing by

an absorbance result with the value of
"zero". This is not mathematically
permissible.

Examples: Calculation of a factor at
one-point calibration; calculation of a
260/280 ratio with nucleic acid
measurements.

Check the reagents and samples used
and repeat the measurement.

Calculation not
possible because of
division by zero.
Absorbance result
or parameter
formula

b is zero.

• An absorbance result was divided by

a "zero" value during the evaluation
of a

Division type method (Dual

wavelength method group). This is
not mathematically permissible.

Check the reagents and samples used
and repeat the measurement.

Do not enter "zero" as a value for the
Formula

b parameter.

Problem

Cause

Solution