3.7.2
Pulse Train Modulation
Define pulse train sequences by using the pulse train editor (
).
Select Mode as “Pulse Trains”.
The editor becomes available when mode is set to pulse trains. There are two input options:
Hexadecimal string or select/unselect boxes of the binary representation. The maximal sequence
length is 256 bits or 64 (hexadecimal) chars when using the GUI. Via direct remote programming up to
4096 bits can be programmed.
The pulse width and pulse period can be set in ranges of milliseconds or microseconds and selected
by the button next to the digits (
Figure 3-k Editor for Pulse Trains