16.

ERRORS

Err2

- Value beyond the zero range

Err3

- Value beyond the tare range

Err4

- Calibration mass or start mass beyond the acceptable

range (

±1% for weight, ±10 for start mass).

Err5

- Mass of a single piece lower than the scale division.

Err7

- Power down time was to short (should be over 3s)

Err8

- Operation time exceeded (tarring, zeroing, start mass

estimation, adjustment)

null

- Zero value from the AD converter

FULL2

- Measurement range overflow

LH

- Start mass error, the mass on the weighing platform

is beyond the acceptable range (from -5% to +15% of
start mass)

5–FULL

- Display range overflow in totalizing


NOTICE:

1. The user is informed about occurrence of errors Err2, Err3, Err4,

Err5, Err8, null by means of a short sound signal (1 second).

2. Error FULL2 is signalled by a sound signal lasting until the over

excessive load is taken off the pan.


17.

RESPECTIVE STANDARDS


PUE C/31H/EX indicator is manufactured in accordance with the following
standards:

- EN 61326-1:2006 Electrical equipment for measurement, control and
laboratory use - EMC requirements -- Part 1: General requirements,

- EN 61010-1:2011 Safety requirements for electrical equipment for
measurement, control, and laboratory use -- Part 1: General requirements,

- EN 60079-0:2013 + A11:2014-03 Explosive atmospheres -- Part 0:
Equipment – General requirements,

- EN 60079-11:2012 Explosive atmospheres -- Part 11: Equipment
protection by intrinsic safety "i".