Errors, Respective standards – RADWAG PUE C indicator User Manual
Page 45

EXI-01-04-06-14-GB
- 45 -
16.
ERRORS
Err2
- Value beyond the zero range
Err3
- Value beyond the tare range
Err4
- Calibration mass or start mass beyond the acceptable
range (
±1% for weight, ±10 for start mass).
Err5
- Mass of a single piece lower than the scale division.
Err7
- Power down time was to short (should be over 3s)
Err8
- Operation time exceeded (tarring, zeroing, start mass
estimation, adjustment)
null
- Zero value from the AD converter
FULL2
- Measurement range overflow
LH
- Start mass error, the mass on the weighing platform
is beyond the acceptable range (from -5% to +15% of
start mass)
5–FULL
- Display range overflow in totalizing
NOTICE:
1. The user is informed about occurrence of errors Err2, Err3, Err4,
Err5, Err8, null by means of a short sound signal (1 second).
2. Error FULL2 is signalled by a sound signal lasting until the over
excessive load is taken off the pan.
17.
RESPECTIVE STANDARDS
PUE C/31H/EX indicator is manufactured in accordance with the following
standards:
- EN 61326-1:2006 Electrical equipment for measurement, control and
laboratory use - EMC requirements -- Part 1: General requirements,
- EN 61010-1:2011 Safety requirements for electrical equipment for
measurement, control, and laboratory use -- Part 1: General requirements,
- EN 60079-0:2013 + A11:2014-03 Explosive atmospheres -- Part 0:
Equipment – General requirements,
- EN 60079-11:2012 Explosive atmospheres -- Part 11: Equipment
protection by intrinsic safety "i".