EXI-01-04-06-14-GB

- 43 -

• Take off the load, message is displayed for 1 sec., the scale

shows the adjustment submenu name:

• Adjustment process may be interrupted at any time by pressing

key, the following message informs about abortion of a process:


NOTICE:

1. It must be remembered to perform calibration when the pan is

unloaded!

2. If the adjustment takes more than 15 sec. the program displays an

error message , short sound signal is heard. Press

key and

perform the adjustment again simultaneously maintaining as stable
ambient conditions as possible!


Save changes and return to weighing:

See point - 11.2. – return to weighing.


15.2. Start mass estimation

If the scale does not require the full calibration process, or if the user lacks
demanded number of mass standards, it is possible to adjust only a new
start mass.

Procedure:

• Following point 11 of this manual enter submenu, next: